Related Experiment Video
Updated: Mar 16, 2026

Peptides from Phage Display Library Modulate Gene Expression in Mesenchymal Cells and Potentiate Osteogenesis in Unicortical Bone Defects
Published on: December 10, 2010
Does PEEK/HA Enhance Bone Formation Compared With PEEK in a Sheep Cervical Fusion Model?
William R Walsh1, Matthew H Pelletier2, Nicky Bertollo2
1Surgical & Orthopaedic Research Laboratories, Prince of Wales Clinical School, Prince of Wales Hospital, UNSW Australia, Level 1 Clinical Sciences Building, Randwick, NSW, 2031, Australia. wrwalshau@gmail.com.
Background:
Polyetheretherketone (PEEK) has a wide range of clinical applications but does not directly bond to bone. Bulk incorporation of osteoconductive materials including hydroxyapatite (HA) into the PEEK matrix is a potential solution to address the formation of a fibrous tissue layer between PEEK and bone and has not been tested.
Questions/Purposes:
Using in vivo ovine animal models, we asked: (1) Does PEEK-HA improve cortical and cancellous bone ongrowth compared with PEEK? (2) Does PEEK-HA improve bone ongrowth and fusion outcome in a more challenging functional ovine cervical fusion model?
Methods:
The in vivo responses of PEEK-HA Enhanced and PEEK-OPTIMA® Natural were evaluated for bone ongrowth in the form of dowels implanted in the cancellous and cortical bone of adult sheep and examined at 4 and 12 weeks as well as interbody cervical fusion at 6, 12, and 26 weeks. The bone-implant interface was evaluated with radiographic and histologic endpoints for a qualitative assessment of direct bone contact of an intervening fibrous tissue later. Gamma-irradiated cortical allograft cages were evaluated as well.
Results:
Incorporating HA into the PEEK matrix resulted in more direct bone apposition as opposed to the fibrous tissue interface with PEEK alone in the bone ongrowth as well as interbody cervical fusions. No adverse reactions were found at the implant-bone interface for either material. Radiography and histology revealed resorption and fracture of the allograft devices in vivo.
Conclusions:
Incorporating HA into PEEK provides a more favorable environment than PEEK alone for bone ongrowth. Cervical fusion was improved with PEEK-HA compared with PEEK alone as well as allograft bone interbody devices.
Clinical Relevance:
Improving the bone-implant interface with a PEEK device by incorporating HA may improve interbody fusion results and requires further clinical studies.
Related Concept Videos
Bone Formation by Intramembranous Ossification
The process begins when mesenchymal cells in the embryonic skeleton gather together and differentiate into osteogenic cells, which then develop into ...
Bone Formation by Endochondral Ossification
Bone Remodeling