Related Experiment Video
Updated: Mar 16, 2026

Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Poor Reliability of Wrist Blood Pressure Self-Measurement at Home: A Population-Based Study
Edoardo Casiglia1, Valérie Tikhonoff2, Federica Albertini2
1From the Department of Medicine, University of Padua, Italy. edoardo.casiglia@unipd.it.
Abstract:
The reliability of blood pressure measurement with wrist devices, which has not previously been assessed under real-life circumstances in general population, is dependent on correct positioning of the wrist device at heart level. We determined whether an error was present when blood pressure was self-measured at the wrist in 721 unselected subjects from the general population. After training, blood pressure was measured in the office and self-measured at home with an upper-arm device (the UA-767 Plus) and a wrist device (the UB-542, not provided with a position sensor). The upper-arm-wrist blood pressure difference detected in the office was used as the reference measurement. The discrepancy between office and home differences was the home measurement error. In the office, systolic blood pressure was 2.5% lower at wrist than at arm (P=0.002), whereas at home, systolic and diastolic blood pressures were higher at wrist than at arm (+5.6% and +5.4%, respectively; P<0.0001 for both); 621 subjects had home measurement error of at least ±5 mm Hg and 455 of at least ±10 mm Hg (bad measurers). In multivariable linear regression, a lower cognitive pattern independently determined both the systolic and the diastolic home measurement error and a longer forearm the systolic error only. This was confirmed by logistic regression having bad measurers as dependent variable. The use of wrist devices for home self-measurement, therefore, leads to frequent detection of falsely elevated blood pressure values likely because of a poor memory and rendition of the instructions, leading to the wrong position of the wrist.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Measurement of Blood Pressure
Errors occurring during blood pressure monitoring
Several factors...
Pre-Procedural Guidelines for Assessing Blood Pressure