Flexible optical intubation via the Ambu Aura-i vs blind intubation via the single-use LMA Fastrach: a prospective
Carlos A Artime1, Alfonso Altamirano1, Katherine C Normand1
1Department of Anesthesiology, UTHealth, The University of Texas Health Science Center at Houston, McGovern Medical School, Houston, TX, USA.
Study Objective:
This study was designed to compare the Ambu Aura-i to the single-use LMA Fastrach regarding time to intubation, success rate, and airway morbidity in patients undergoing elective surgery requiring general anesthesia.
Design:
Prospective, randomized controlled trial.
Setting:
Academic medical center.
Patients:
Sixty-five adult patients scheduled for elective surgery requiring general anesthesia.
Interventions:
Patients were randomized into 2 groups. Group A (n=33) were intubated using Ambu Aura-i and the Ambu aScope 2, a disposable flexible intubating scope, whereas those in group B (n=33) were blindly intubated using the Intubating Laryngeal Mask Airway (ILMA).
Measurements:
First-attempt intubation success rate, overall intubation success rate, time to intubation, incidence of airway morbidity.
Main Results:
The data demonstrated that time for endotracheal intubation in the ILMA group was significantly shorter than in the Ambu Aura-i group (P<.05). There was no difference in the first-attempt intubation success rate (Aura-i=26/33, 78.8%; ILMA=27/33, 81.8%; P=.757) or the overall intubation success rate (Aura-i=29/33, 87.9%; ILMA=31/33, 93.9%; P=.392) between the groups. Four patients (12%) in the Ambu Aura-i group had a failed intubation; 1 was due to a failure of the aScope monitor, whereas 3 were due to inability to visualize the glottis. Two patients (7%) in the ILMA group had a failed intubation due to esophageal intubation. There was no statistically significant difference in airway morbidity between the 2 groups.
Conclusions:
The data suggest that intubation with the ILMA is faster but that first-attempt and overall intubation success rates were comparable in both groups. The results suggest that although the flexible intubating scope-guided Aura-i does not outperform blind intubation via the ILMA, the technique is comparable in terms of first-attempt and overall intubation success rate.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....