Related Experiment Video
Updated: Mar 15, 2026

Direct Pressure Monitoring Accurately Predicts Pulmonary Vein Occlusion During Cryoballoon Ablation
Published on: February 26, 2013
Factors Influencing Diaphragmatic Compound Motor Action Potentials During Cryoballoon Ablation for Atrial
Parikshit S Sharma1, Santosh K Padala1, J Jenkins Thompson1
1Division of Cardiology, Virginia Commonwealth University Medical Center, Richmond, Virginia, USA.
Introduction:
The diaphragmatic compound motor action potentials (CMAPs) have been used to predict and prevent phrenic nerve injury (PNI) during cryoballoon ablation of right pulmonary veins. We sought to assess factors that influence the amplitude of the surface CMAP recordings.
Methods And Results:
We analyzed CMAPs from consecutive patients undergoing cryoballoon ablation for paroxysmal atrial fibrillation. CMAP recordings were obtained using electrocardiography electrodes positioned in the "modified lead I" method while stimulating the right PN, until loss of capture (ascertained by palpation and fluoroscopy of the right hemi-diaphragm). A total of 55 patients (age 63 ± 11 years; 60% men; body mass index [BMI] 31 ± 6) had adequate CMAP recordings and were included for evaluation of CMAP signals. CMAPs demonstrated 2 distinct components, an early higher amplitude signal (pacing artifact) and a later lower amplitude signal (true diaphragmatic CMAP). There was no significant change in the true CMAP recording amplitude with decrease in stimulus strength (P = 0.1). There was no impact of BMI on CMAP amplitude (P = 0.93). There was a significant phasic respiratory variation in CMAP amplitude with a mean decrease in CMAP amplitude of 10.8% (range: 8-12%) with inspiration lasting an average of 2 beats (P < 0.001). A decrease in CMAP amplitude of >30% was noted in 6 cases (11%) and termination of cryoablation prevented PNI.
Conclusion:
Diaphragmatic CMAP amplitude is not affected by stimulus strength or BMI. There is a significant respirophasic decrease in CMAP signal amplitude with inspiration. It is important to be aware of this variation to avoid premature termination of cryoablation.
More Related Videos
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015
28:13Catheter Ablation in Combination With Left Atrial Appendage Closure for Atrial Fibrillation
Published on: February 26, 2013
Related Concept Videos
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials