Prospective Comparison of Liver Stiffness Measurements between Two Point Shear Wave Elastography Methods: Virtual
Hyunsuk Yoo1, Jeong Min Lee2, Jeong Hee Yoon1
1Department of Radiology, Seoul National University Hospital, Seoul 03080, Korea.
Korean Journal of Radiology
|September 3, 2016
Summary
Virtual Touch Quantification (VTQ) and Elastography Point Quantification (ElastPQ) show comparable success rates and reliable liver stiffness measurements. However, these point shear wave elastography methods are not interchangeable due to significant agreement limits.
Area of Science:
- Medical Imaging
- Hepatology
- Diagnostic Ultrasound
Background:
- Liver stiffness (LS) measurement is crucial for assessing liver fibrosis.
- Point shear wave elastography (pSWE) offers non-invasive methods for LS assessment.
- Virtual Touch Quantification (VTQ) and Elastography Point Quantification (ElastPQ) are two pSWE techniques.
Purpose of the Study:
- To prospectively compare the technical success rate and reliability of VTQ and ElastPQ.
- To correlate liver stiffness measurements obtained by VTQ and ElastPQ.
- To evaluate the intraobserver reproducibility of ElastPQ.
Main Methods:
- Prospective comparison of VTQ and ElastPQ in 85 patients with chronic liver disease.
- Statistical analysis included chi-squared tests, Wilcoxon signed-rank test, Spearman correlation, and Bland-Altman analysis.
- Intraobserver reproducibility of ElastPQ assessed using Bland-Altman limits and intraclass correlation coefficient (ICC).
Main Results:
- Similar technical success rates (98.8% for VTQ vs. 95.3% for ElastPQ) and reliable measurements (95.3% for VTQ vs. 90.6% for ElastPQ).
- No significant difference in mean LS measurements between VTQ (1.71 ± 0.47 m/s) and ElastPQ (1.66 ± 0.41 m/s).
- Strong correlation (r = 0.820) between LS measurements, but a large 95% limit of agreement (27.5% of the mean). High intraobserver reproducibility for ElastPQ (ICC = 0.991).
Conclusions:
- VTQ and ElastPQ demonstrate comparable technical success and reliability for liver stiffness assessment.
- While LS measurements are strongly correlated, the methods are not interchangeable due to a wide limit of agreement.
- ElastPQ exhibits excellent intraobserver reproducibility.