Rapid Scan Electron Paramagnetic Resonance Opens New Avenues for Imaging Physiologically Important Parameters In Vivo
Published on: September 26, 2016
Comparison of Continuous Wave and Rapid Scan X-band Electron Paramagnetic Resonance of Irradiated Clipped Fingernails
Hanan Elajaili1, Joseph McPeak1, Alexander Romanyukha2
1Department of Chemistry and Biochemistry, University of Denver, Denver, CO 80208 USA.
Abstract:
X-band rapid scan electron paramagnetic resonance (EPR) measures the free radicals in irradiated clipped fingernails with higher signal-to-noise (S/N) and lower standard deviation of the signal amplitude for replicate measurements than does conventional continuous wave (CW) EPR in the same measurement time. For a clipped fingernail sample irradiated to 10 Gy and data acquisition time of 30 s with B1 = 8.5 μT, the S/N for rapid scan is >2000 for the absorption spectrum and 1200 for the corresponding first derivative. The standard deviation for replicate measurements of signal amplitude is ~1%. For CW spectra, the S/N depends on the modulation amplitude. The CW signal has a first derivative peak-to-peak linewidth of 0.95 mT. For 30 s of CW acquisition time, the S/N was 30 for a conservative modulation amplitude of 0.2 mT and B1 of 2.3 μT or 90 for a modulation amplitude of 0.4 mT and B1 of 3.2 μT, which resulted in standard deviations of replicate measurements of 5% or 2%, respectively.