Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
12.6K
Tibial component with and without stem extension in a trabecular metal cone construct
Marrigje F Meijer1, Alexander L Boerboom2, Martin Stevens2
1Department of Orthopaedics, University Medical Center Groningen, University of Groningen, PO Box 30.001, 9700 RB, Groningen, The Netherlands. M.F.Meijer@umcg.nl.
Summary
This study found that tibial components with or without a stem following trabecular metal cone reconstruction show similar stability and strain distribution. Bone mineral density minimally impacted anterior/posterior tilt, suggesting stems may not be mandatory for tibial plateau reconstruction.
Area of Science:
- Orthopedic Surgery
- Biomedical Engineering
- Materials Science
Background:
- Trabecular metal cones are used to reconstruct major bone defects in total knee arthroplasty.
- The necessity of stem extension for tibial components after trabecular metal cone reconstruction remains unclear.
- Stem implantation can enhance stability but may carry potential negative side effects.
Purpose of the Study:
- To evaluate the stability and strain distribution of tibial plateau reconstructions using trabecular metal cones with and without stem extension.
- To determine if prosthetic stability is influenced by bone mineral density.
- To compare biomechanical outcomes of stemmed versus non-stemmed tibial components in this context.
Main Methods:
- Seven pairs of cadaveric tibiae underwent tibial revision arthroplasty for 2B bone defects, with one tibia per pair receiving a stemmed component and the other a non-stemmed component.
- Specimens were subjected to one bodyweight load, alternating between medial and lateral tibial components.
- Implant-bone micro-motions, bone strains, and bone mineral density were measured and analyzed.
Main Results:
- No significant difference in strain distribution was observed between stemmed and non-stemmed components.
- Non-stemmed components exhibited a slight, clinically insignificant increase in varus tilt (0.14°).
- Bone mineral density significantly correlated with anterior/posterior tilt (ρ: -0.72).
Conclusions:
- Tibial components implanted with or without a stem after trabecular metal cone reconstruction yield comparable biomechanical conditions regarding stability and strain distribution.
- The findings suggest that stem extension may not be mandatory in these cases.
- Further in vivo studies are recommended to confirm these biomechanical observations and guide clinical decision-making.