Related Experiment Video
Updated: Mar 15, 2026

Author Spotlight: High-Quality Quantum Dot Nanobeads for Sensitive Fluorescent Lateral Flow Immunoassays
Published on: June 28, 2024
Detection of norovirus virus-like particles using a surface plasmon resonance-assisted fluoroimmunosensor optimized
Hiroki Ashiba1, Yuki Sugiyama2, Xiaomin Wang3
1Electronics and Photonics Research Institute, National Institute of Advanced Industrial Science and Technology (AIST), Tsukuba, Japan.
Abstract:
A highly sensitive biosensor to detect norovirus in environment is desired to prevent the spread of infection. In this study, we investigated a design of surface plasmon resonance (SPR)-assisted fluoroimmunosensor to increase its sensitivity and performed detection of norovirus virus-like particles (VLPs). A quantum dot fluorescent dye was employed because of its large Stokes shift. The sensor design was optimized for the CdSe-ZnS-based quantum dots. The optimal design was applied to a simple SPR-assisted fluoroimmunosensor that uses a sensor chip equipped with a V-shaped trench. Excitation efficiency of the quantum dots, degree of electric field enhancement by SPR, and intensity of autofluorescence of a substrate of the sensor chip were theoretically and experimentally evaluated to maximize the signal-to-noise ratio. As the result, an excitation wavelength of 390nm was selected to excite SPR on an Al film of the sensor chip. The sandwich assay of norovirus VLPs was performed using the designed sensor. Minimum detectable concentration of 0.01ng/mL, which corresponds to 100 virus-like particles included in the detection region of the V-trench, was demonstrated.