Related Experiment Video
Updated: Mar 15, 2026

Experimental Methods for Trapping Ions Using Microfabricated Surface Ion Traps
Published on: August 17, 2017
Successive Resonances for Ion Ejection at Arbitrary Frequencies in an Ion Trap
Dalton T Snyder1, R Graham Cooks2
1Department of Chemistry and Center for Analytical Instrumentation Development, Purdue University, West Lafayette, IN, 47907, USA.
Abstract:
The use of successive resonances for ion ejection is demonstrated here as a method of scanning quadrupole ion traps with improvement in both resolution and sensitivity compared with single frequency resonance ejection. The conventional single frequency resonance ejection waveform is replaced with a dual-frequency waveform. The two included frequencies are spaced very closely and their relative amplitudes are adjusted so that the first frequency that ions encounter excites them to higher amplitudes where space charge effects are less prominent, thereby giving faster and more efficient ejection when the ions come into resonance with the second frequency. The method is applicable at any arbitrary frequency, unlike double and triple resonance methods. However, like double and triple resonance ejection, ejection using successive resonances requires the rf and AC waveforms to be phase-locked in order to retain mass accuracy and mass precision. The improved performance is seen in mass spectra acquired by rf amplitude scans (resonance ejection) as well as by secular frequency scans. Graphical Abstract ᅟ.
Related Concept Videos
Mass Analyzers: Common Types
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
Sound Waves: Resonance
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Mass Spectrum: Interpretation
Series Resonance

