Related Experiment Video
Updated: Mar 15, 2026

Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
Automated oxygen titration and weaning with FreeO2 in patients with acute exacerbation of COPD: a pilot randomized
François Lellouche1, Pierre-Alexandre Bouchard1, Maude Roberge1
1Research Centre.
Introduction:
We developed a device (FreeO2) that automatically adjusts the oxygen flow rates based on patients' needs, in order to limit hyperoxia and hypoxemia and to automatically wean them from oxygen.
Objective:
The aim of this study was to evaluate the feasibility of using FreeO2 in patients hospitalized in the respiratory ward for an acute exacerbation of COPD.
Methods:
We conducted a randomized controlled trial comparing FreeO2 vs manual oxygen titration in the respiratory ward of a university hospital. We measured the perception of appropriateness of oxygen titration and monitoring in both groups by nurses and attending physicians using a Likert scale. We evaluated the time in the target range of oxygen saturation (SpO2) as defined for each patient by the attending physician, the time with severe desaturation (SpO2 <85%), and the time with hyperoxia (SpO2 >5% above the target). We also recorded length of stay, intensive care unit admissions, and readmission rate. Fifty patients were randomized (25 patients in both groups; mean age: 72±8 years; mean forced expiratory volume in 1 second: 1.00±0.49 L; and mean initial O2 flow 2.0±1.0 L/min).
Results:
Nurses and attending physicians felt that oxygen titration and monitoring were equally appropriate with both O2 administration systems. The percentage of time within the SpO2 target was significantly higher with FreeO2, and the time with severe desaturation and hyperoxia was significantly reduced with FreeO2. Time from study inclusion to hospital discharge was 5.8±4.4 days with FreeO2 and 8.4±6.0 days with usual oxygen administration (P=0.051).
Conclusion:
FreeO2 was deemed as an appropriate oxygen administration system by nurses and physicians of a respiratory unit. This system maintained SpO2 at the target level better than did manual titration and reduced periods of desaturation and hyperoxia. Our results also suggest that FreeO2 has the potential to reduce the hospital length of stay.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue,...
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...
COPD: Management Using Bronchodilators and Corticosteroids