Randomized Cross-Over Study Comparing Two Infusion Sets for CSII in Daily Life
Guido Freckmann1, Stephan Arndt2, Albrecht Fießelmann3
11 Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Ulm, Germany.
Continuous subcutaneous insulin infusion (CSII) therapy uses insulin infusion sets (IIS). This study found two soft-cannula IIS were safe and effective for type 1 diabetes management, with no significant difference in insertion pain.
Area of Science:
- Endocrinology
- Medical Devices
- Diabetes Management
Background:
- Continuous subcutaneous insulin infusion (CSII) is crucial for type 1 diabetes.
- Insulin infusion sets (IIS) are key components of CSII systems.
- Limited research exists on IIS performance and safety.
Purpose of the Study:
- To compare the performance and safety of two soft-cannula insulin infusion sets (IIS).
- To evaluate user experience and adverse events associated with two specific IIS models.
Main Methods:
- An 8-week, open-label, randomized, crossover, multicenter study.
- 80 type 1 diabetes patients on CSII evaluated Accu-Chek® FlexLink Plus (FL Plus) and Accu-Chek® FlexLink (FL) IIS for 4 weeks each.
- Data collected via patient diaries and visual analogue scale (VAS) for insertion pain.
Main Results:
- No significant difference in insertion pain between FL Plus (VAS 6.1 mm) and FL (VAS 5.8 mm).
- Unplanned changes occurred in 20.0% (FL Plus) and 16.0% (FL) of IIS changes.
- 13 adverse events (2 serious) occurred over 4243 patient days, with some related to IIS.
- 18.7% of cannulas kinked; overall patient satisfaction was high for both IIS.
Conclusions:
- Both evaluated IIS are safe and effective for CSII therapy.
- There is no discernible difference in insertion pain between the two IIS.
- Both IIS demonstrate acceptable safety and efficacy profiles for managing type 1 diabetes.
Related Concept Videos
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
Crossover Experiments
Crossover designs are performed even with smaller sample sizes since the samples can act as their controls. These are better than simple randomized trials since patients are exposed to all the treatments.
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
Bioequivalence Experimental Study Designs: Repeated Measures, Cross-Over, Carry-Over, and Latin Square Designs
Bioequivalence Experimental Study Designs: Completely Randomized and Randomized Block Designs
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...