A Protocol of Manual Tests to Measure Sensation and Pain in Humans
Published on: December 19, 2016
Reliability of Pain Measurements Using Computerized Cuff Algometry: A DoloCuff Reliability and Agreement Study.
Jack Kvistgaard Olsen1, Dilay Kesgin Fener1, Eva Elisabet Waehrens1,2
1Department of Rheumatology, The Parker Institute, Copenhagen University Hospital, Bispebjerg and Frederiksberg, Frederiksberg, Denmark.
The DoloCuff device demonstrates excellent inter- and intrarater reliability for assessing pressure pain threshold, pressure pain tolerance, and temporal summation index in healthy individuals. This new pain assessment method is reliable for clinical use.
Area of Science:
- Pain research
- Algometry
- Biomedical engineering
Background:
- Computerized pneumatic cuff pressure algometry (CPA) is a novel method for pain assessment.
- The reliability of the DoloCuff device for CPA has not been previously established.
- Establishing reliability is crucial for validating new pain measurement tools.
Purpose of the Study:
- To evaluate the inter-rater and intra-rater reliability of the DoloCuff for measuring pressure pain parameters.
- To assess the reliability of pressure pain threshold (PT), pressure pain tolerance (PTol), and temporal summation index (TSI) using CPA.
- To determine the clinical utility of the DoloCuff in healthy subjects.
Main Methods:
- Twenty healthy adults (20-29 years) participated in the study.
- Two trained raters assessed pressure pain parameters three times at 24-hour intervals.
- Reliability was analyzed using intraclass correlation coefficients (ICC) and Bland-Altman plots.
Main Results:
- Excellent inter-rater reliability was found for PT (ICC=0.88), PTol (ICC=0.86), and TSI (ICC=0.81).
- Excellent intra-rater reliability was observed for PT (ICC=0.81) and PTol (ICC=0.89).
- TSI showed borderline excellent/good intra-rater reliability (ICC=0.75).
Conclusions:
- The DoloCuff device provides reliable measures for PT, PTol, and TSI in healthy individuals.
- The CPA method using DoloCuff is a dependable tool for pain assessment.
- The findings support the use of DoloCuff for reliable pain parameter measurement in research and clinical settings.
More Related Videos
08:40Isokinetic Robotic Device to Improve Test-Retest and Inter-Rater Reliability for Stretch Reflex Measurements in Stroke Patients with Spasticity
Published on: June 12, 2019
07:50Measuring Local Tissue Strains in Tendons via Open-Source Digital Image Correlation
Published on: January 27, 2023
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: