Related Experiment Video
Updated: Mar 15, 2026

Author Spotlight: In-Depth Morphometric Examination and Quantification of Native Lens Structure Using Whole Mount Imaging
Published on: January 19, 2024
Microscopic analysis of an opacified OFT CRYL® hydrophilic acrylic intraocular lens
Bruna Vieira Ventura1,2, Kyle Douglas MacLean3, Wagner Lira1
1Fundação Altino Ventura (FAV) - Recife, PE, Brazil.
Abstract:
A 51-year-old patient underwent posterior vitrectomy with perfluoropropane gas injection, phacoemulsification, and implantation of an Oft Cryl® hydrophilic acrylic intraocular lens (IOL) because of traumatic retinal detachment and cataract in the right eye. On the first postoperative day, gas was filling the anterior chamber because of patient's non-compliance in terms of head positioning, and was reabsorbed within one week. Eight months later, the patient returned complaining of a significant decrease in vision. IOL opacification was noticed by slit-lamp examination. The lens was explanted to undergo gross and light microscopic analysis. The lens was also stained with the alizarin red method for calcium identification. Light microscopic analysis confirmed the presence of granular deposits, densely distributed in an overall circular pattern in the central part of the lens optic. The granules stained positive for calcium. This is the first case of the opacification of this type of hydrophilic lens. Surgeons should be aware of this potential postoperative complication, and the use of hydrophilic IOLs should be avoided in procedures involving intracameral gas because of the risk of IOL opacification.