Reliability of different temporary myocardial pacing leads
S Wirtz1, H D Schulte, J Winter
1Department of Thoracic and Cardiovascular Surgery, Heinrich Heine University of Düsseldorf, FRG.
The Thoracic and Cardiovascular Surgeon
|June 1, 1989
Summary
This study compared temporary myocardial pacing leads for reliability after open-heart surgery. Bipolar pacing leads demonstrated superior performance, especially for atrial sensing, making them recommended for postoperative pacing.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Devices
Background:
- Temporary myocardial pacing leads are crucial for managing postoperative arrhythmias after open-heart surgery.
- Assessing the reliability of different pacing electrodes is essential for patient care.
Purpose of the Study:
- To compare the reliability of five different temporary myocardial pacing electrodes.
- To evaluate pacing and sensing functions over a 10-day postoperative period.
Main Methods:
- A standardized implantation technique was used for 147 ventricular and 81 atrial electrodes in 149 patients.
- Stimulation threshold, P- and R-wave sensing, and impedance were measured daily using a pacing system analyzer.
- Statistical analysis of mean values and comparison of pacing/sensing failures were performed.
Main Results:
- Bipolar pacing leads (Osypka TME 64a/66a) showed higher reliability than unipolar wires (Ethicon HD 12S, Osypka TME 60/60a), particularly in atrial sensing.
- Medtronic 6400/6500 leads exhibited good reliability for ventricular pacing, suitable for VVI-mode pacing.
- Parameter trends and failure rates were analyzed to determine lead performance.
Conclusions:
- Specially developed bipolar temporary pacemaker electrodes are recommended for reliable atrial and ventricular pacing post-surgery.
- Bipolar leads offer improved reliability for temporary myocardial pacing up to 10 days postoperatively.
- Selection of appropriate pacing leads impacts the efficacy of managing postoperative arrhythmias.
Related Concept Videos
Electrocardiogram Fundamentals
Introduction
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Dysrhythmias V: Evaluating Dysrhythmias
Dysrhythmias, also known as arrhythmias, are disturbances in the heart's rhythm that range from benign to life-threatening. A thorough evaluation is crucial for appropriate management and involves a comprehensive medical history, physical examination, and various diagnostic tests.Medical HistorySymptoms: Collect detailed information on palpitations, dizziness, syncope, chest pain, and fatigue. Note their onset, frequency, and triggers.Previous Cardiac Issues: Document any history of heart...