Related Experiment Video
Updated: Mar 15, 2026

Author Spotlight: Developing a Disposable Dosator for Preclinical Testing of Dry Powder Inhalers in Small Animal Models
Published on: August 18, 2023
Variability in Delivered Dose from Pressurized Metered-Dose Inhaler Formulations Due to a Delay Between Shake and
Ross H M Hatley1, Jacob Parker1, John N Pritchard1
11 Philips Respironics Respiratory Drug Delivery (UK) Ltd, a business of Philips Electronics UK Limited, Chichester, United Kingdom .
Background:
Pressurized metered-dose inhalers (pMDIs) should be shaken before use to prevent creaming or sedimentation of the drugs in solution; however, data published on this topic are limited, and it is rarely specified how soon after shaking the device should be actuated. Delays between shaking and firing the pMDI have previously been shown to cause significant inhomogeneity in delivered dose. We studied the effect of various shake-fire delays on the drug delivered from five commercially available pMDIs commonly prescribed for asthma and chronic obstructive pulmonary disease to assess the potential variability in delivered dose.
Methods:
The pMDI formulations tested were the Flovent HFA, Ventolin Evohaler, Airomir Inhaler, and Symbicort (suspension pMDIs), and the QVAR 100 Inhaler (solution pMDI). Each pMDI was shaken for 5 seconds before attachment to a dosage unit sampling apparatus collection tube and filter, and it was actuated once with shake-fire delays of 0, 5, 10, 20, 30, 40, 50, and 60 seconds. Analysis of the eluates from the collection tubes and filters was performed by using high-performance liquid chromatography. Three of each pMDI were tested twice with each time delay.
Results:
All of the suspension pMDIs produced variable amounts of drug over the shake-fire delays tested. A comparison of the delivered doses after the 0- and 60-second delays showed that the drug delivered increased for the Flovent HFA (320%), Ventolin Evohaler (346%), and Airomir Inhaler (230%) pMDIs; decreased for the Symbicort budesonide (75%) and formoterol fumarate (76%) pMDI; and remained consistent for the QVAR 100 Inhaler pMDI.
Conclusions:
The amount of drug delivered can vary widely over different shake-fire delays with suspension pMDIs. Therefore, guidance should be given to users/caregivers on the timing of firing after shaking their device, particularly with pediatrics, who may take time to become receptive to accepting their medication after pMDI shaking and before dose administration.
Related Concept Videos
Inhaled Medications
Dosage Regimen: Individualization
Modified-Release Drug Delivery Systems: Influencing Factors
Pharmaceutical Alternatives: Polymorphic Form-Related and Particle Size-Related Therapeutic Nonequivalence
Modified-Release Drug Delivery Systems: Drug Release Characteristics
Modified-Release Drug Delivery Systems: Rate-Programmed I