A prospective multicenter clinical feasibility study of a new automatic speaking valve for postlaryngectomy voice rehabilitation
Related Experiment Video
Updated: Mar 15, 2026

Identification and Protection of the Recurrent Laryngeal Nerve during Transoral Robotic Thyroidectomy
Published on: October 24, 2025
L Lansaat1, B J de Kleijn2, F J M Hilgers3,4
1Department of Head and Neck Oncology and Surgery, Netherlands Cancer Institute, Plesmanlaan 121, 1066 CX, Amsterdam, The Netherlands.
Abstract:
Evaluation of short- and long-term clinical feasibility and exploration of limitations and advantages of a new automatic speaking valve (ASV) for laryngectomized patients with integrated HME, the Provox FreeHands FlexiVoice (FlexiVoice). This ASV not only enables automatic, but also manual closure of the valve. A multicenter, prospective clinical study in 40 laryngectomized patients was conducted. Participants were asked to use the FlexiVoice for 26 weeks. The primary outcome measure was long-term compliance. Secondary outcome measures were: patient preference, hours of FlexiVoice use, device life of adhesive, voice and speech quality, and quality of life. After 26 weeks, 15 patients (37.5 %) were using the FlexiVoice on a daily basis, for a mean of 12.64 h/day (SD ± 5.03). Ten patients (25 %) were using the device on a non-daily basis, for a mean of 3.76 h/day (SD ± 2.07). The remaining 15 patients (37.5 %) discontinued using the FlexiVoice. Sixty percent of the 25 long-term users applied both automatic and manual closure of the valve. Unpredictable fixation of the adhesive was the main reason for discontinuing or not using the FlexiVoice on a daily basis. Overall, 18 patients (45 %) preferred the FlexiVoice, 16 patients (40 %) their usual HME, 3 patients (7.5 %) their usual ASV, 1 patient (2.5 %) preferred no device at all, and in 2 patients preference was not recorded. The minor technical issues identified could be corrected. The Provox FreeHands FlexiVoice appears to be a useful ASV, which allows for hands-free speech in a larger proportion of laryngectomized patients in the present cohort. The additional manual closure option of the device is beneficial for maintaining the adhesive seal longer.
Related Concept Videos
Tracheostomy Suctioning II: Procedure
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...