Lactate Sensors on Flexible Substrates
Related Experiment Video
Updated: Mar 14, 2026

Hollow Microneedle-based Sensor for Multiplexed Transdermal Electrochemical Sensing
Published on: June 1, 2012
Xuesong Yang1, Timothy Fu2, Pavan Kumar Kota3
1Electrical Engineering, University of Texas‒Arlington, Arlington, TX 76019, USA. xuesong.yang@mavs.uta.edu.
Abstract:
Lactate detection by an in situ sensor is of great need in clinical medicine, food processing, and athletic performance monitoring. In this paper, a flexible, easy to fabricate, and low-cost biosensor base on lactate oxidase is presented. The fabrication processes, including metal deposition, sol-gel IrOx deposition, and drop-dry enzyme loading method, are described in detail. The loaded enzyme was examined by scanning electron microscopy. Cyclic voltammetry was used to characterize the sensors. Durability, sensibility, and selectivity of the biosensors were examined. The comparison for different electrode sizes and different sensing film materials was conducted. The sensor could last for four weeks with an average surface area normalized sensitivity of 950 nA/(cm² mM) and 9250 nA/(cm² mM) for Au-based electrodes, and IrOx-modified electrodes respectively, both with an electrode size of 100 × 50 μm. The self-referencing method to record noises simultaneously with the working electrode greatly improved sensor sensitivity and selectivity. The sensor showed little response to interference chemicals, such as glutamate and dopamine.
More Related Videos
13:42Dry Film Photoresist-based Electrochemical Microfluidic Biosensor Platform: Device Fabrication, On-chip Assay Preparation, and System Operation
Published on: September 19, 2017
11:18Manufacturing of a Nafion-coated, Reduced Graphene Oxide/Polyaniline Chemiresistive Sensor to Monitor pH in Real-time During Microbial Fermentation
Published on: January 7, 2019