Related Experiment Video
Updated: Mar 14, 2026

Noninvasive In Vivo Small Animal MRI and MRS: Basic Experimental Procedures
Published on: October 20, 2009
In vivo MR imaging with simultaneous RF transmission and reception
Sung-Min Sohn1, J Thomas Vaughan1, Russell L Lagore1
1Center for Magnetic Resonance Research, Minneapolis, Minnesota, USA.
Purpose:
To present a practical scheme of a simultaneous radiofrequency (RF) transmit (Tx) and receive (Rx) (STAR) system for MRI, discuss the challenges and solutions, and show preliminary in vivo MR images obtained with this new technique.
Methods:
A remotely controlled STAR system was built and tested with a transverse electromagnetic head coil on a 4T (Oxford, 90 cm-bore) MRI scanner equipped with an Agilent DirectDrive console (Agilent, Santa Clara, CA). In vivo head images have been acquired using continuous sweep excitation and acquisition.
Results:
The bench test and MR experimental results show our STAR system to have high isolation (60 dB) between Tx and Rx, with insensitivity to load swings created by head motion. To acquire in vivo head images, ultralow RF peak power of 50 mW was used.
Conclusion:
A novel motion-insensitive STAR MRI technique was developed and experimentally tested. The first in vivo MR images using this method were acquired. Magn Reson Med 76:1932-1938, 2016. © 2016 International Society for Magnetic Resonance in Medicine.