Fast Megavoltage Computed Tomography: A Rapid Imaging Method for Total Body or Marrow Irradiation in Helical
Taiki Magome1, Akihiro Haga2, Yutaka Takahashi3
1Department of Radiological Sciences, Faculty of Health Sciences, Komazawa University, Tokyo, Japan; Department of Radiology, The University of Tokyo Hospital, Tokyo, Japan; Masonic Cancer Center, University of Minnesota, Minneapolis, Minnesota.
Purpose:
Megavoltage computed tomographic (MVCT) imaging has been widely used for the 3-dimensional (3-D) setup of patients treated with helical tomotherapy (HT). One drawback of MVCT is its very long imaging time, the result of slow couch speeds of approximately 1 mm/s, which can be difficult for the patient to tolerate. We sought to develop an MVCT imaging method allowing faster couch speeds and to assess its accuracy for image guidance for HT.
Methods And Materials:
Three cadavers were scanned 4 times with couch speeds of 1, 2, 3, and 4 mm/s. The resulting MVCT images were reconstructed using an iterative reconstruction (IR) algorithm with a penalty term of total variation and with a conventional filtered back projection (FBP) algorithm. The MVCT images were registered with kilovoltage CT images, and the registration errors from the 2 reconstruction algorithms were compared. This fast MVCT imaging was tested in 3 cases of total marrow irradiation as a clinical trial.
Results:
The 3-D registration errors of the MVCT images reconstructed with the IR algorithm were smaller than the errors of images reconstructed with the FBP algorithm at fast couch speeds (2, 3, 4 mm/s). The scan time and imaging dose at a speed of 4 mm/s were reduced to 30% of those from a conventional coarse mode scan. For the patient imaging, faster MVCT (3 mm/s couch speed) scanning reduced the imaging time and still generated images useful for anatomic registration.
Conclusions:
Fast MVCT with the IR algorithm is clinically feasible for large 3-D target localization, which may reduce the overall time for the treatment procedure. This technique may also be useful for calculating daily dose distributions or organ motion analyses in HT treatment over a wide area. Automated integration of this imaging is at least needed to further assess its clinical benefits.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...