Related Experiment Video
Updated: Mar 14, 2026

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Comparison of the hypothetical (57)Co brachytherapy source with the (192)Ir source
Mohammad Taghi Bahreyni Toossi1, Mahdi Ghorbani1, Atefeh Rostami1
1Biomedical Engineering and Medical Physics Department, Faculty of Medicine, Shahid Beheshti University of Medical Sciences, Tehran, Iran.
Aim Of The Study:
The (57)Co radioisotope has recently been proposed as a hypothetical brachytherapy source due to its high specific activity, appropriate half-life (272 days) and medium energy photons (114.17 keV on average). In this study, Task Group No. 43 dosimetric parameters were calculated and reported for a hypothetical (57)Co source.
Material And Methods:
A hypothetical (57)Co source was simulated in MCNPX, consisting of an active cylinder with 3.5 mm length and 0.6 mm radius encapsulated in a stainless steel capsule. Three photon energies were utilized (136 keV [10.68%], 122 keV [85.60%], 14 keV [9.16%]) for the (57)Co source. Air kerma strength, dose rate constant, radial dose function, anisotropy function, and isodose curves for the source were calculated and compared to the corresponding data for a (192)Ir source.
Results:
The results are presented as tables and figures. Air kerma strength per 1 mCi activity for the (57)Co source was 0.46 cGyh(-1) cm 2 mCi(-1). The dose rate constant for the (57)Co source was determined to be 1.215 cGyh(-1)U(-1). The radial dose function for the (57)Co source has an increasing trend due to multiple scattering of low energy photons. The anisotropy function for the (57)Co source at various distances from the source is more isotropic than the (192)Ir source.
Conclusions:
The (57)Co source has advantages over (192)Ir due to its lower energy photons, longer half-life, higher dose rate constant and more isotropic anisotropic function. However, the (192)Ir source has a higher initial air kerma strength and more uniform radial dose function. These properties make (57)Co a suitable source for use in brachytherapy applications.
Related Concept Videos
Biological Effects of Radiation
Nuclear Transmutation
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Radiation: Applications
The average...
Absorption of Radiation