Related Experiment Video
Updated: Mar 14, 2026

An Immunocompetent Murine Model for Laser Interstitial Thermal Therapy of Glioblastoma
Published on: November 15, 2024
The use of custom 3D printed stereotactic frames for laser interstitial thermal ablation: technical note
Nicholas J Brandmeir1, James McInerney1, Brad E Zacharia1
1Department of Neurosurgery, Penn State Health Milton S. Hershey Medical Center, Hershey, Pennsylvania.
A novel 3D printed stereotactic frame for laser interstitial thermotherapy (LITT) offers a faster, more accurate approach to treating brain lesions. This innovative method simplifies surgery and may become a preferred technique for LITT procedures.
Area of Science:
- Neurosurgery
- Medical Device Technology
- 3D Printing Applications
Background:
- Laser interstitial thermotherapy (LITT) is increasingly used for cranial lesions.
- Existing LITT techniques have varied benefits and drawbacks regarding fiber placement.
- There is a need for improved stereotactic methods in LITT.
Purpose of the Study:
- To introduce and evaluate a novel, customized 3D printed stereotactic frame for LITT procedures.
- To assess the feasibility, accuracy, and efficiency of this new system.
- To compare this method with existing stereotactic techniques for LITT.
Main Methods:
- A customized 3D printed stereotactic frame was designed using preoperative imaging and skull fiducials.
- The frame was attached to implanted fiducials on the day of surgery for LITT catheter placement.
- The system was utilized in 5 consecutive patients, with 2 also undergoing needle biopsy.
Main Results:
- Intraoperative and postoperative imaging confirmed accurate LITT catheter placement and lesion creation.
- Mean operating room time was 45 minutes (26 minutes excluding biopsy cases).
- The STarFix microTargeting system facilitated rapid procedures with extensive planning and no intraoperative fixation.
Conclusions:
- The customized 3D printed stereotactic frame represents a significant advancement for LITT.
- This system offers advantages such as reduced operating time and enhanced accuracy.
- It may become a preferred stereotactic method for LITT, particularly in MRI-suite settings.