Differences in visual quality with orientation of a rotationally asymmetric bifocal intraocular lens design
Aiswaryah Radhakrishnan1, Carlos Dorronsoro1, Susana Marcos1
1From the Laboratory of Visual Optics and Biophotonics, Instituto de Óptica Daza de Valdés, Consejo Superior de Investigaciones Científicas, Madrid, Spain.
Purpose:
To evaluate visual and perceptual performance for different orientations of a rotationally asymmetric bifocal intraocular lens (IOL) (M-Plus) simulated optically using a simultaneous vision simulator.
Setting:
Instituto de Optica, Madrid, Spain.
Design:
Prospective observational study.
Methods:
Perceptual quality and decimal high-contrast visual acuity (HCVA) was measured under cycloplegia for 8 orientations of the asymmetric bifocal IOL phase pattern at far, intermediate, and near distances simulated with a simultaneous vision simulator using face images and tumbling E targets. The preferred orientation at each distance was calculated as the centroid of the data for 8 orientations. The visual Strehl value was calculated using the subjects' ocular aberrations and multifocal pattern at each orientation. Optical predictions were obtained by implementing a differential visual Strehl values-based ideal observer model.
Results:
The study comprised 20 subjects (aged 21 to 62 years). Horizontal orientation (near segment at 0 or 180 degrees ± 45 [SD]) was preferred by 14 subjects and by 13 subjects at far and near distances, respectively; 8 subjects showed strong orientation preferences. The mean difference in preferred orientation between far and near was 27 ± 22 degrees. No significant differences in HCVA were observed. Optical predictions correlated strongly and significantly with measurements (far r = 0.71, near r = 0.62; P < .0001). The mean difference between measurement and simulation in the preferred orientation was 28 ± 29 degrees at far and 36 ± 28 degrees at near.
Conclusions:
The perception varied for different orientations of an asymmetric bifocal IOL design tested using a simultaneous vision simulator. Optimum orientation was driven by interactions of the design with the eye's optical aberrations.
Financial Disclosure:
None of the authors has a financial or proprietary interest in any material or method mentioned.
More Related Videos
06:55Scanning Light Scattering Profiler SLPS Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
05:14Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Related Concept Videos
Properties of Enantiomers and Optical Activity
Focusing of Light in the Eye
Rotation of Asymmetric Top
The relationship between the angular momentum of any rigid body and its angular velocity, both of which are vectors, involves the moment of inertia. The moment of inertia is a scalar quantity only for spherically symmetric...
Unsymmetric Bending - Angle of Neutral Axis
When a bending moment is applied at an angle θ concerning the vertical axis of a symmetrical member, it can be resolved into components along the member's principal...
Accessory Structures of the Eye
Angle Closure Glaucoma: Treatment