The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
3.7K
Novel, Patient-Specific Instruments for Acetabular Preparation and Cup Placement
Ran Schwarzkopf1, Erik Schnaser2, Taiki Nozaki3
1Department of Orthopaedic Surgery, NYU Langone Medical Center Hospital for Joint Diseases, New York, New York.
Surgical Technology International
|October 12, 2016
Summary
Patient-specific instrumentation for total hip arthroplasty (THA) demonstrated reproducible acetabular component placement. This novel approach using Bullseye Hip Replacement Instruments® shows accurate anteversion, abduction, and component sizing in cadaveric models.
Area of Science:
- Orthopedic Surgery
- Biomedical Engineering
Background:
- Patient-specific instrumentation (PSI) offers potential benefits in joint replacements, including improved accuracy and efficiency.
- In total hip arthroplasty (THA), PSI aims to enhance surgical precision, implant longevity, and patient outcomes.
- A novel PSI system was developed to improve acetabular bone resection and cup placement in THA.
Purpose of the Study:
- To evaluate the efficacy of a novel patient-specific instrumentation (PSI) system for total hip arthroplasty (THA).
- To assess the accuracy of acetabular bone resection and component placement using the Bullseye Hip Replacement Instruments®.
Main Methods:
- A cadaveric study involving 14 acetabuli was conducted.
- Pre-operative CT/MRI scans informed the use of Bullseye Hip Replacement Instruments®.
- Two experienced hip surgeons performed THA using the PSI technique, with post-operative CT scans for analysis.
Main Results:
- The Bullseye Hip Replacement Instruments® were used to implant 14 acetabuli.
- Post-operative CT analysis revealed an average acetabular cup anteversion of 15.50°.
- Average acetabular cup abduction/inclination was 35.9°, with accurate component sizing matching pre-operative plans.
Conclusions:
- The Bullseye Hip Replacement Instruments® provide reproducible acetabular component placement in terms of anteversion and abduction.
- This PSI system demonstrates accurate sizing of the acetabular component in THA.
- The findings suggest potential for improved surgical accuracy in total hip arthroplasty using this novel instrumentation.