Related Experiment Video
Updated: Mar 13, 2026

Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
In phantom assessment of superficial doses under TomoTherapy irradiation
Margherita Zani1, Cinzia Talamonti1, Marta Bucciolini1
1Dipartimento di Scienze Biomediche, Sperimentali e Cliniche, Università degli Studi di Firenze, Italy; Azienda Ospedaliero Universitaria Careggi, Firenze, Italy.
Purpose:
Aim of this work is the assessment of build-up and superficial doses of different clinical Head&Neck plans delivered with Helical TomoTherapy (HT) (Accuray, Sunnyvale, CA). Depth dose profiles and superficial dose points were measured in order to evaluate the Treatment Planning System (TPS) capability of an accurate dose modeling in regions of disequilibrium. Geometries and scattering conditions were investigated, similar to the ones generally encountered in clinical treatments.
Methods:
Measurements were performed with two dosimeters: Gafchromic® EBT3 films (Ashland Inc., Wayne, NJ) and a synthetic single crystal diamond detector (PTW-Frieburg microDiamond, MD). A modified version of the Alderson RANDO phantom was employed to house the detectors. A comparison with TPS data was carried out in terms of dose difference (DD) and distance-to-agreement (DTA).
Results:
DD between calculated data and MD measurements are within 4% even in points with high spatial dose variation. For depth profiles, EBT3 data show a DDmax of 3.3% and DTAmax of 2.2mm, in low and high gradient regions, respectively, and compare well with MD data. EBT3 superficial points always results in measured doses lower than TPS evaluated ones, with a maximum DTA value of 1.5mm.
Conclusions:
Doses measured with the two devices are in good agreement and compare well with calculated data. The deviations found in the present work are within the reference tolerance level, suggesting that the HT TPS is capable of a precise dose estimation both in superficial regions and in correspondence with interfaces between air and PMMA.
More Related Videos
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022