Novel thin layer flow-cell screen-printed graphene electrode for enzymatic sensors
Hussein Kanso1, María Begoña González García1, Laura Fernández Llano1
1DropSens, S.L., Edificio CEEI, Parque Tecnológico de Asturias, 33428 Llanera, Asturias, Spain.
Abstract:
A new Screen-printed electrodes (SPE) integrated in one channel flow-cell was developed. The one channel flow-cell is attached and directly changeable with electrode. In the new flow-cell the injection is done through an "in-line luer injection port" which can be less aggressive than wall-jet flow cell for a biological recognition element immobilized on the surface of the electrode. The sample volume can be easily controlled by the operator through a syringe. In this novel thin layer flow-cell screen-printed electrodes, the working electrode was modified with graphene materials, and an enhancement of electroactive area to 388% over a standard electrode was found. This new configuration was applied to study the entrapped cellobiose dehydrogenase from the ascomycete Corynascus thermophilus (CtCDH) in a photocrosslinkable PVA-based polymer. The calibration curve of lactose using optimized parameters shows a wide linear measurement ranges between 0.25 and 5mM. A good operational stability of the CtCDH-PVA-modified graphene electrode is obtained, which keeps the same initial activity during 8h and exhibits a good storage stability with a decrease of only 9% in analytical response after 3 months storage at 4◦C.