Effect of Body Mass Index on Digital Templating for Total Hip Arthroplasty
Robert A Sershon1, Alejandro Diaz1, Daniel D Bohl1
1Department of Orthopedic Surgery, Rush University Medical Center, Chicago, Illinois.
Background:
Digital templating is becoming more prevalent in orthopedics. Recent investigations report high accuracy using digital templating in total hip arthroplasty (THA); however, the effect of body mass index (BMI) on templating accuracy is not well described.
Methods:
Digital radiographs of 603 consecutive patients (645 hips) undergoing primary THA by a single surgeon were digitally templated using OrthoView (Jacksonville, FL). A 25-mm metallic sphere was used as a calibration marker. Preoperative digital hip templates were compared with the final implant size. Hips were stratified into groups based on BMI: BMI <30 (315), BMI 30-35 (132), BMI 35-40 (97), and BMI >40 (101).
Results:
Accuracy between templating and final size did not vary by BMI for acetabular or femoral components. Digital templating was within 2 sizes of the final acetabular and femoral implants in 99.1% and 97.1% of cases, respectively.
Conclusion:
Digital templating is an effective means of predicting the final size of THA components. BMI does not appear to play a major role in altering THA digital templating accuracy.