A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Long-Term Stability and Clinical Utility of Amplified Atrial Electrograms in a Single-Lead ICD System with Floating
Nicole E Worden1, Musab Alqasrawi1, Alexander Mazur1
1Division of Cardiovascular Medicine, University of Iowa Hospitals and Clinics, University of Iowa Carver College of Medicine, Iowa City, Iowa.
Background:
Available atrial electrograms in implantable cardioverter defibrillators (ICDs) improve arrhythmia diagnosis, allow monitoring for atrial fibrillation, and may reduce the risk of inappropriate therapies. A recently introduced ICD system using a single-lead with floating atrial electrodes provides diagnostic capability of a dual-chamber system without placing an additional lead. Data on long-term clinical performance of this system are limited.
Methods:
We retrospectively analyzed data from 35 consecutive patients implanted with Biotronik VR-T DX devices and LinoxSmart DX leads. (Biotronik, SE & Co., Berlin, Germany) RESULTS: Of 35 patients (77% male, age 52 ± 11.28 years), 32 were followed for a mean of 432 ± 197 days (range 56-765). During implantation, average preamplified and amplified sinus P-wave amplitudes were 2.61 ± 1.39 mV (range 0.9-6.8 mV) and 8.7 ± 4.51 mV (range 1.4-18 mV), respectively. Despite statistically significant variations, the amplified P-wave amplitude measurements (calculated mean values over 3 months) remained within a clinically acceptable range during follow-up (5.4-8.7 mV). R-wave amplitude and ventricular pacing threshold measurements were stable over time. A total of 13 stored arrhythmia events (three ventricular tachycardia, eight supraventricular tachycardia, two atrial fibrillation) were reviewed. All of them showed readily interpretable atrial electrograms. Eight out of 10 (80%) supraventricular events were correctly classified by the device. Three patients received inappropriate ICD therapies.
Conclusion:
The single-lead ICD system using a floating atrial dipole provides reliable recording of atrial signals during sinus rhythm and arrhythmias. Our data suggest that the system may offer diagnostic advantages of a dual-chamber device without potential risks of an additional atrial lead.
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...