Interventional dual-energy imaging-Feasibility of rapid kV-switching on a C-arm CT system
1Radiological Sciences Lab, Stanford University, Stanford, California 94305.
Purpose:
In the last years, dual-energy CT imaging has shown clinical value, thanks to its ability to differentiate materials based on their atomic number and to exploit different properties of images acquired at two different energies. C-arm CT systems are used to guide procedures in the interventional suite. Until now, there are no commercially available systems that employ dual-energy material decomposition. This paper explores the feasibility of implementing a fast kV-switching technique on a clinically available angiographic system for acquiring dual-energy C-arm CT images.
Methods:
As an initial proof of concept, a fast kV-switching approach was implemented on an angiographic C-arm system and the peak tube voltage during 3D rotational scans was measured. The tube voltage measurements during fast kV-switching scans were compared to corresponding measurements on kV-constant scans. Additionally, to prove stability of the requested exposure parameters, the accuracy of the delivered tube current and pulse width were also recorded and compared. In a first phantom experiment, the voxel intensity values of the individual tube voltage components of the fast kV-switching scans were compared to their corresponding kV-constant scans. The same phantom was used for a simple material decomposition between different iodine concentrations and pure water using a fast kV-switching protocol of 81 and 125 kV. In the last experiment, the same kV-switching protocol as in the phantom scan was used in an in vivo pig study to demonstrate the clinical feasibility.
Results:
During rapid kV-switching acquisitions, the measured tube voltage of the x-ray tube during fast switching scans has an absolute deviation of 0.23 ± 0.13 kV compared to the measured tube voltage produced during kV-constant acquisitions. The stability of the peak tube voltage over different scan requests was about 0.10 kV for the low and 0.46 for the high energy kV-switching scans and less than 0.1 kV for kV-constant scans, indicating slightly lower stability for kV-switching scans. The tube current resulted in a relative deviation of -1.6% for the low and 6.6% overestimation for the high tube voltage of the kV-switching scans compared to the kV-constant scans. The pulse width showed no deviation for the longer pulse width and only minor deviations (0.02 ± 0.02 ms) for the shorter pulse widths compared to the kV-constant scans. The phantom experiment using different iodine concentrations showed an accurate correlation (R2 > 0.99) between the extracted intensity values in the kV-switching and kV-constant reconstructed volumes, and allows for an automatic differentiation between contrast concentration down to 10% (350 mg/ml iodine) and pure water under low-noise conditions. Preliminary results of iodine and soft tissue separation showed also promising results in the first in vivo pig study.
Conclusions:
The feasibility of dual-energy imaging using a fast kV-switching method on an angiographic C-arm CT system was investigated. Direct measurements of beam quality in the x-ray field demonstrate the stability of the kV-switching method. Phantom and in vivo experiments showed that images did not deviate from those of corresponding kV-constant scans. All performed experiments confirmed the capability of performing fast kV-switching scans on a clinically available C-arm CT system. More complex material decomposition tasks and postprocessing steps will be part of future investigations.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT