Plastic strains during stent deployment have a critical influence on the rate of corrosion in absorbable magnesium stents
Related Experiment Video
Updated: Mar 13, 2026

Ferromagnetic Bare Metal Stent for Endothelial Cell Capture and Retention
Published on: September 18, 2015
Emmet Galvin1, Christy Cummins2, Shoichiro Yoshihara3
1School of Mechanical and Manufacturing Engineering, Dublin City University, Dublin 9, Ireland. emmet.galvin3@mail.dcu.ie.
Abstract:
Magnesium stents are a promising candidate in the emerging field of absorbable metallic stents (AMSs). In this study, the mechanical and corrosion performance of dog-bone specimens and a specific stent design of a magnesium alloy, WE43, are assessed experimentally in terms of their corrosion behaviour and mechanical integrity. It is shown that plastic strains that are induced in the struts of the stent during stent deployment have a critical influence in directing subsequent corrosion behaviour within the material. In addition, the deployment and scaffolding characteristics of the magnesium stent are elucidated and contrasted with those of a commercial stainless steel stent. The magnesium stent is found to support higher levels of cyclic strain amplitude than the stainless steel stent, even prior to degradation, and this may play a role in reducing in-stent restenosis. This study provides new insights into the experimental performance of a current AMS design and material whilst demonstrating the critical influence of plastic strain on the corrosion performance and scaffolding ability of an AMS.
Related Concept Videos
Temperature Dependent Deformation
Residual Stresses
Plastic Deformations
Plastic Deformations
Normal Strain under Axial Loading
Stress-Strain Diagram - Ductile Materials