Related Experiment Video
Updated: Mar 13, 2026

Three-Dimensional Reconstruction of Orbital Fractures
Published on: May 16, 2025
CT of facial fracture fixation: an experimental study of artefact reducing methods
Elina M Peltola1, Teemu Mäkelä1, Ville Haapamäki1
11 Department of Radiology, HUS Medical Imaging Center, Radiology, University of Helsinki and Helsinki University Hospital, Helsinki, Finland.
Objectives:
This study aimed to determine the optimal post-operative CT imaging method that enables best visualization of facial bony structures in the vicinity of osteosynthesis material.
Methods:
Conducted at Töölö Hospital (Helsinki, Finland), this study relied on scanning a phantom with CBCT, 64-slice CT and high-definition multislice CT with dual-energy scan (providing monochromatic images of 70-, 100-, 120- and 140-keV energy levels) and iterative reconstruction (IR) methods. Two radiologists assessed the image quality, and the assessments were analyzed. In addition, a physicist performed a semi-quantitative analysis of the metal-induced artefacts.
Results:
The three subjects most easily assessed were the loose screw and both the bone structure and the fracture further away from the screw and the plate. Soft tissues adjacent to the screw and the plate remained more difficult for assessment. Both image interpreters agreed that the artefacts disturbed their assessments under dual energy. Metal artefacts disturbed the least under multislice CT with IR [adaptive statistical iterative reconstruction (ASiR) and VEO]. Neither interpreter found metal suppression helpful in CBCT.
Conclusions:
CBCT with or without a metal artefact reduction algorithm was not optimal for post-operative facial imaging compared with multislice CT with IR. Multislice CT with ASiR filtering offered good image quality performance with fast image volume reconstruction, representing the current sweet spot in post-operative maxillofacial imaging.
More Related Videos
15:11Surgical Fixation of Sternal Fractures: Preoperative Planning and a Safe Surgical Technique Using Locked Titanium Plates and Depth Limited Drilling
Published on: January 5, 2015
04:13Less-Invasive Technique for Non-stabilized Mandibular Fracture in Mouse Models
Published on: September 27, 2024