An improved device for bioimpedance deviation measurements based on 4-electrode half bridge.
Yang Liu1, Xiaoyan Qiao2, Gang Li1
1School of Precision Instrument and Opto-Electronics Engineering, Tianjin University, 300072 Tianjin, China.
The Review of Scientific Instruments
|November 3, 2016
Summary
This study presents an improved device for real-time bioimpedance deviation detection, crucial for monitoring body fluid changes. The new system achieves high resolution and a large dynamic range, enabling precise measurements.
Area of Science:
- Biomedical Engineering
- Physiological Monitoring
Background:
- Monitoring body fluid changes is critical in healthcare.
- Real-time bioimpedance deviation detection offers a promising approach for this monitoring.
- Existing methods face challenges with accuracy and dynamic range.
Purpose of the Study:
- To develop an improved device for sensitive bioimpedance deviation detection.
- To enhance the accuracy and dynamic range of body fluid change monitoring systems.
- To overcome limitations of current bioimpedance measurement techniques.
Main Methods:
- Utilized a self-balancing half bridge circuit to deduct static impedance baselines.
- Employed the 4-electrode technique to eliminate contact impedance interference.
- Implemented a digital lock-in algorithm with oversampling to boost effective resolution.
Main Results:
- The developed device can detect minimal bioimpedance deviations as low as 1 Ω.
- Effective measurement achieved even with a high bioimpedance baseline of 10 kΩ.
- The system demonstrated high resolution, magnification, large dynamic range, and self-adaptability.
Conclusions:
- The proposed device significantly advances real-time bioimpedance deviation detection capabilities.
- This technology holds potential for improved monitoring of body fluid changes.
- The system's performance highlights its suitability for various physiological monitoring applications.
Related Concept Videos
Potentiometry: Membrane Electrodes
2.1K
Membrane electrodes, also known as p-ion electrodes, use membranes that selectively interact with free analyte ions, generating a potential difference across the membrane. The resulting membrane potential, known as the asymmetry potential, is not zero even when analyte concentrations on both sides of the membrane are equal. The membrane's response is typically not selective to a single analyte but proportional to the concentration of all ions in the sample solution capable of interacting at...
2.1K
Wheatstone Bridge
1.3K
An ohmmeter is a resistance-measuring device. It works by applying a voltage to a resistor of unknown resistance and measuring the current across the resistor. The resistance value is deduced using Ohm's law. Usually, the standard configuration of an ohmmeter comprises a voltmeter or an ammeter. However, such configurations are limited in accuracy because the meters alter the voltage applied to the resistor and the current that flows through it.
Thus, for accurate resistance measurements, a...
Thus, for accurate resistance measurements, a...
1.3K