Related Experiment Video
Updated: Mar 12, 2026

High-throughput Quantitative Real-time RT-PCR Assay for Determining Expression Profiles of Types I and III Interferon Subtypes
Published on: March 24, 2015
Results of the NIST National Ball Plate Round Robin
G W Caskey1, S D Phillips1, B R Borchardt1
1National Institute of Standards and Technology, Gaithersburg, MD 20899-0001.
Abstract:
This report examines the results of the ball plate round robin administered by NIST. The round robin was part of an effort to assess the current state of industry practices for measurements made using coordinate measuring machines. Measurements of a two-dimensional ball plate (240 mm by 240 mm) on 41 coordinate measuring machines were collected and analyzed. Typically, the deviations of the reported X and Y coordinates from the calibrated values were within ± 5 μm, with some coordinate deviations exceeding 20.0 μm. One of the most significant observations from these data was that over 75 % of the participants failed to correctly estimate their measurement error on one or more of the ball plate spheres.