Doppler Optical Coherence Tomography of Retinal Circulation
Published on: September 18, 2012
Optical coherence tomography: influence of contrast concentration on image quality and diagnostic confidence
Florian Blachutzik1, Stephan Achenbach2, Holger Nef3
1Department of Cardiology, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), University Hospital Erlangen, Ulmenweg 18, 91054, Erlangen, Germany. florian.blachutzik@uk-erlangen.de.
Abstract:
OCT requires intracoronary injection of contrast agent to remove blood from the coronary lumen during data acquisition, which is a possible limitation of this method. Aim of this study was to analyze the influence of iodine concentration on image quality and diagnostic certainty of optical coherence tomography (OCT). OCT sequences acquired using contrast agent with a reduced concentration of 150 mg iodine/ml and a standard concentration of 350 mg iodine/ml were analyzed. Cross-sectional images with a spacing of 10 mm were evaluated regarding image quality and diagnostic confidence. A total of 67 OCT sequences acquired in 24 patients were analyzed. 31 sequences were acquired using contrast agent with a concentration of 150 mg iodine/ml and 36 sequences with a concentration of 350 mg iodine/ml. The percentage of remaining blood streaks in the cross sections was significantly lower for 350 mg iodine/ml compared to 150 mg iodine/ml (19 ± 21 vs. 34 ± 26%, p = 0.013). Contrast with 350 mg iodine/ml showed a significantly higher percentage of completely flushed pullback length as compared to 150 mg iodine/ml (78 ± 24 vs. 58 ± 27%, p = 0.004). Diagnostic certainty was significantly higher for 350 mg iodine/ml than for 150 mg iodine/ml (Likert scale average 1.4 ± 0.7 vs. 2.1 ± 1.2, p < 0.001; Likert scale: 1 = absolutely confident, 2 = confident with slight doubts, 3 = doubtful/not confident, 4 = non-diagnostic). Regarding image quality and diagnostic certainty, contrast agent with a concentration of 350 mg iodine/ml is superior to 150 mg iodine/ml.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Imaging Studies for Cardiovascular System V: CT
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...