Objectives:
To achieve the efficient usage of contrast material (CM) in high-pitch CT aortography, an appropriate duration of the CM injection is crucial. We used a modification of the double-level test bolus method for determination of proper injection duration with the aim of evaluating the image quality of tailored-duration CM injection compared with that of a fixed duration.
Materials And Methods:
The institutional review board approved retrospective review of 80 consecutive subjects who had undergone high-pitch 70-kVp CT aortography with a modified double-level test bolus method. The interval between peak enhancement at the aortic root and femoral artery was derived from the time/attenuation curves. A total of 40 subjects underwent CT aortography with individually set duration time from the results. The remaining subjects underwent CT aortography with a fixed-duration time. The density values at several parts of the aorta were assessed. The differences in image quality and CM amount used for each method were assessed by Welch test.
Results:
The injection duration was almost 50% shorter (median, 15 seconds; range, 11-25 seconds) when individually tailored. The mean CM amount was reduced by 50% (46.2-23.9 mL, P < 0.01). The range of mean CT attenuation throughout the aorta was not significantly different between the 2 methods (316-327 HU and 305-321 HU, P > 0.05, respectively).
Conclusions:
The modified double-level test bolus method in high-pitch CT aortography can significantly reduce the amount of CM without adversely affecting image quality.
More Related Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies VII: Vascular Imaging