Reference Clinical Database for Fixation Stability Metrics in Normal Subjects Measured with the MAIA Microperimeter
Marco U Morales1, Saker Saker1, Craig Wilde1
1Division of Clinical Neurosciences, University of Nottingham, Queen's Medical Centre, Nottingham, UK.
Purpose:
The purpose of this study was to establish a normal reference database for fixation stability measured with the bivariate contour ellipse area (BCEA) in the Macular Integrity Assessment (MAIA) microperimeter.
Methods:
Subjects were 358 healthy volunteers who had the MAIA examination. Fixation stability was assessed using two BCEA fixation indices (63% and 95% proportional values) and the percentage of fixation points within 1° and 2° from the fovea (P1 and P2). Statistical analysis was performed with linear regression and Pearson's product moment correlation coefficient.
Results:
Average areas of 0.80 deg2 (min = 0.03, max = 3.90, SD = 0.68) for the index BCEA@63% and 2.40 deg2 (min = 0.20, max = 11.70, SD = 2.04) for the index BCEA@95% were found. The average values of P1 and P2 were 95% (min = 76, max = 100, SD = 5.31) and 99% (min = 91, max = 100, SD = 1.42), respectively. The Pearson's product moment test showed an almost perfect correlation index, r = 0.999, between BCEA@63% and BCEA@95%. Index P1 showed a very strong correlation with BCEA@63%, r = -0.924, as well as with BCEA@95%, r = -0.925. Index P2 demonstrated a slightly lower correlation with both BCEA@63% and BCEA@95%, r = -0.874 and -0.875, respectively.
Conclusions:
The single parameter of the BCEA@95% may be taken as accurately reporting fixation stability and serves as a reference database of normal subjects with a cutoff area of 2.40 ± 2.04 deg2 in MAIA microperimeter.
Translational Relevance:
Fixation stability can be measured with different indices. This study originates reference fixation values for the MAIA using a single fixation index.