Diagnosis of Intoxication by the Organophosphate VX: Comparison Between an Electrochemical Sensor and Ellman´s
Miroslav Pohanka1, Martina Hrabinova2, Kamil Kuca3
1Centre of Advanced Studies and Department of Toxicology, Faculty of Military Health Sciences, University of Defense / Trebesska 1575, 50001 Hradec Kralove, Czech Republic. rau@atlas.cz.
Abstract:
An electrochemical sensor is introduced as a tool applicable for diagnosis of intoxication by cholinesterase inhibitors caused by the well-known nerve agent VX. The traditional Ellman method was chosen for comparison with the sensor's analytical parameters. Both methods are based on estimation of blood cholinesterase inhibition as a marker of intoxication. While Ellman´s method provided a limit of detection of 5.2´10-7 M for blood containing VX, the electrochemical sensor was able to detect 4.0´10-7 M. Good correlation between both methods was observed (R = 0.92). The electrochemical sensor could be considered a convenient tool for a fast yet accurate method, easily available for field as well as laboratory use. Time and cost savings are key features of the sensor-based assay.