Effect of diagnostic cone-beam computed tomography protocols on image quality, patient dose, and lesion detection
Jijo Paul1, Annamma Chacko2, Paola Saccomandi3
1Department of Diagnostic and Interventional Radiology, University Hospital, Johann Wolfgang Goethe University, Theodor-Stern-Kai 7, 60590 Frankfurt, Germany.
Objective:
To evaluate the effect of cone-beam computed tomography (CBCT) image acquisition protocols on image quality, lesion detection, delineation, and patient dose.
Methods:
100-patients and a CTDI phantom combined with an electron density phantom were examined using four different CBCT-image acquisition protocols during image-guided transarterial chemoembolization (TACE). Protocol-1 (time: 6s, tube rotation: 360°), protocol-2 (5s, 300°), protocol-3 (4s, 240°) and protocol-4 (3s, 180°) were used. The protocols were first investigated using a phantom. The protocols that were found to be clinically appropriate in terms of image quality and radiation dose were then assessed on patients. A higher radiation dose and/or a poor image quality were inappropriate for the patient imaging. Patient dose (patient-entrance dose and dose-area product), image quality (Hounsfield Unit, noise, signal-to-noise ratio and contrast-to-noise ratio), and lesion delineation (tumor-liver contrast) were assessed and compared using appropriate statistical tests. Lesion detectability, sensitivity, and predictive values were estimated for CBCT-image data using pre-treatment patient magnetic resonance imaging.
Results:
The estimated patient dose showed no statistical significance (p>0.05) between protocols-2 and -3; the assessed image quality between these protocols manifested insignificant difference (p>0.05). Two other phantom protocols were not considered for patient imaging due to significantly higher dose (protocols-1) and poor image quality (protocol-4). Lesion delineation and detection were insignificant (p>0.05) between protocols-2 and -3. Lesion sensitivities generated were 81-89% (protocol-2) and 81-85% (protocol-3) for different lesion types.
Conclusion:
Data acquisition using protocols-2 and -3 provided good image quality, lesion detection and delineation with acceptable patient dose during CBCT-imaging mainly due to similar frame numbers acquired.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
11:38Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
X-ray Imaging