Related Experiment Video
Updated: Jul 29, 2026

Combustion Characterization and Model Fuel Development for Micro-tubular Flame-assisted Fuel Cells
Published on: October 2, 2016
A recoverable gas-cell diagnostic for the National Ignition Facility
A Ratkiewicz1, L Berzak Hopkins1, D L Bleuel1
1Lawrence Livermore National Laboratory, Livermore, California 95440, USA.
Abstract:
The high-fluence neutron spectrum produced by the National Ignition Facility (NIF) provides an opportunity to measure the activation of materials by fast-spectrum neutrons. A new large-volume gas-cell diagnostic has been designed and qualified to measure the activation of gaseous substances at the NIF. This in-chamber diagnostic is recoverable, reusable and has been successfully fielded. Data from the qualification of the diagnostic have been used to benchmark an Monte Carlo N-Particle Transport Code simulation describing the downscattered neutron spectrum seen by the gas cell. We present early results from the use of this diagnostic to measure the activation of natXe and discuss future work to study the strength of interactions between plasma and nuclei.