The use of routine non density calibrated clinical computed tomography data as a potentially useful screening tool
Christopher John Burke1, Manjiri M Didolkar2, Huiman X Barnhart3
1Hospital for Joint Diseases, NYU Langone Medical Center, Department of Radiology, Division of Musculoskeletal Radiology, NY, NY, USA.
Objectives:
To evaluate whether lumbar vertebral body density CT attenuation values measured in Hounsfield Units (HUs) on routine Computed Tomography (CT) examinations can be reliably measured with limited variability, and to evaluate for a correlation between HUs and bone mineral density as measured by dual energy X-ray absorptiometry (DXA) scan.
Methods:
Retrospective review of a total of 249 routine MDCT examinations, performed to measure HUs at the first non-rib bearing lumbar vertebral body on axial images, cross-referenced to the lateral scout image.
Results:
The overall ICC and RC for intra-reader variability on CT HU were 0.987 (95% CI 0.973 - 0.999) and 15.664 (95% CI 11.66-16.97). The overall ICC and RDC for inter-reader variability on CT HU were 0.952 (95% CI 0.892 - 0.999) and 30.20 (95% CI 23.73 - 34.48). The ICC and RC for interscanner variability were 0.98 (95% CI 0.95 - 0.99) and 16.67 (95% CI 13.13 - 22.85). The correlation between the L1 HUs and L1 BMD, L1 t-score, and overall t-score was 0.437, 0.392, and 0.400, respectively.
Conclusions:
CT attenuation values of the first lumbar vertebra can be measured on routine abdomen CTs with limited variability despite multiple readers and scanners. Correlation between HU and BMD as measured by DXA scan was only weakly positive, and by this method measuring the density of a lumbar vertebral body from a routine MDCT scan does not provide the sensitivity or specificity necessary for a screening test. However above a certain measured value (180 HU), patients have a low chance of osteoporosis and therefore may not need additional screening, potentially limiting radiation exposure and cost.
More Related Videos
07:12Semiautomated Longitudinal Microcomputed Tomography-based Quantitative Structural Analysis of a Nude Rat Osteoporosis-related Vertebral Fracture Model
Published on: September 28, 2017
04:40Identifying Coronary Artery Calcification on Non-gated Computed Tomography Scans
Published on: August 28, 2018
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT