High Mass Ion Detection with Charge Detector Coupled to Rectilinear Ion Trap Mass Spectrometer
Avinash A Patil1, Szu-Wei Chou1,2, Pei-Yu Chang1
1Department of Physics, National Dong Hwa University, Shoufeng, Hualien, Taiwan, 97401, Republic of China.
Abstract:
Conventional linear ion trap mass analyzers (LIT-MS) provide high ion capacity and show their MS n ability; however, the detection of high mass ions is still challenging because LIT-MS with secondary electron detectors (SED) cannot detect high mass ions. To detect high mass ions, we coupled a charge detector (CD) to a rectilinear ion trap mass spectrometer (RIT-MS). Immunoglobulin G ions (m/z ~150,000) are measured successfully with controlled ion kinetic energy. In addition, when mass-to-charge (m/z) ratios of singly charged ions exceed 10 kTh, the detection efficiency of CD is found to be greater than that of SED. The CD can be coupled to LIT-MS to extend the detection mass range and provide the potential to perform MS n of high mass ions inside the ion trap. Graphical Abstract ᅟ.