Alarm safety and oxygen saturation targets in the Vermont Oxford Network iNICQ 2015 collaborative
J I Hagadorn1,2, D W Sink1,2, M E Buus-Frank3,4,5
1Division of Neonatology, Connecticut Children's Medical Center, Hartford, CT, USA.
Objectives:
To assess progress of neonatal intensive care units (NICUs) participating in the Vermont Oxford Network iNICQ 2015: Alarm Safety Collaborative in achieving Joint Commission 2014 alarm safety goals with respect to oximeters, and to compare patient-level oxygen saturation (SpO2) and oximeter alarm data to local policies.
Study Design:
Prospective multicenter audits in February and August 2015 assessed implementation of policies addressing Joint Commission 2014 Alarm Safety goals, and ascertained SpO2 targets, oximeter alarm settings and compliance with policy-specified SpO2 targets and alarms.
Results:
Eighty-six NICUs completed both audits. Of 13 policies addressing mandated goals, median (interquartile range) 8 (5, 9) policies were implemented at audit 1 and 9 (6, 11) at audit 2 (P=0.004). At audit 1, 28 NICUs had implemented ⩾9 policies versus 47 at audit 2. For 794 infants <31 weeks gestation, <36 weeks postmenstrual age, and on supplemental oxygen, median SpO2 target lower limit was 88% (interquartile range 87%, 90%; range 75% to 94%), upper limit 95% (interquartile range 94%, 96%; range 85% to 100%). High oximeter alarm was set according to local policy for 63% of infants, for whom SpO2 >97% was less frequent than when high alarm was not set to policy (10.1% vs 21.5%, P=0.006).
Conclusions:
Participating NICUs showed significant progress between audits in their implementation of Joint Commission Alarm Safety goals for oximeter monitoring. Oximeter high alarm not set per local policy is associated with increased hyperoxemia in preterm infants. Recommendations to standardize oxygen saturation targets for infants at risk for oxygenation-related outcomes have not been widely adopted.
More Related Videos
06:57Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...