Related Experiment Video
Updated: Mar 10, 2026

Magnetic Resonance Imaging of Multiple Sclerosis at 7.0 Tesla
Published on: February 19, 2021
Evaluation of exposure to (ultra) high static magnetic fields during activities around human MRI scanners
Mahsa Fatahi1, Jolanta Karpowicz2, Krzysztof Gryz2
1Department of Biomedical Magnetic Resonance, H65-ZENIT, Otto-von-Guericke-University Magdeburg, Leipziger Street 44, 39120, Magdeburg, Germany. Mahsa.Fatahi@ovgu.de.
Objective:
To assess the individual exposure to the static magnetic field (SMF) and the motion-induced time-varying magnetic field (TVMF) generated by activities in an inhomogeneous SMF near high and ultra-high field magnetic resonance imaging (MRI) scanners. The study provides information on the level of exposure to high and ultra-high field MRI scanners during research activities.
Materials And Methods:
A three-axis Hall magnetometer was used to determine the SMF and TVMF around human 3- and 7-Tesla (T) MRI systems. The 7-T MRI scanner used in this study was passively shielded and the 3-T scanner was actively shielded and both were from the same manufacturer. The results were compared with the exposure restrictions given by the International Commission on Non-Ionizing Radiation Protection (ICNIRP).
Results:
The recorded exposure was highly variable between individuals, although they followed the same instructions for moving near the scanners. Maximum exposure values of B = 2057 mT and dB/dt = 4347 mT/s for the 3-T scanner and B = 2890 mT, dB/dt = 3900 mT/s for 7 T were recorded. No correlation was found between reporting the MRI-related sensory effects and exceeding the reference values.
Conclusions:
According to the results of our single-center study with five subjects, violation of the ICNIRP restrictions for max B in MRI research environments was quite unlikely at 3 and 7 T. Occasions of exceeding the dB/dt limit at 3 and 7 T were almost similar (30% of 60 exposure scenarios) and highly variable among the individuals.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging