Respiratory average CT for attenuation correction in myocardial perfusion SPECT/CT
Duo Zhang1, Bang-Hung Yang2,3, Nien Yun Wu2,3
1Biomedical Imaging Laboratory, Department of Electrical and Computer Engineering, Faculty of Science and Technology, University of Macau, Avenida da Universidade, Taipa, Macau SAR, China.
Cine average CT (CACT) and interpolated average CT (IACT) attenuation correction (AC) improve myocardial perfusion SPECT/CT quantitation compared to helical CT (HCT). IACT further reduces radiation dose, offering a superior alternative for improved imaging accuracy.
Area of Science:
- Nuclear Medicine
- Medical Imaging
- Radiology
Background:
- Attenuation correction (AC) is crucial for accurate SPECT/CT imaging.
- Respiratory motion introduces artifacts in myocardial perfusion SPECT.
- Cine average CT (CACT) and interpolated average CT (IACT) are advanced AC techniques.
Purpose of the Study:
- To evaluate the effectiveness of CACT and IACT for attenuation correction in myocardial perfusion SPECT/CT.
- To compare CACT and IACT with standard helical CT (HCT) based AC.
- To assess the impact of these methods on quantitative accuracy and radiation dose.
Main Methods:
- Computer simulations using the XCAT phantom with varying respiratory motion amplitudes.
- Physical phantom experiments with a clinical SPECT/CT scanner.
- Generation of IACT by averaging interpolated CT phases.
- Reconstruction of SPECT/CT data with AC using CACT, IACT, and HCT.
Main Results:
- CACT and IACT-AC significantly reduced respiratory artifacts and improved quantitative accuracy compared to HCT-AC.
- SPECT reconstructed with CACT showed the highest similarity to the original phantom data (RDmax <8%).
- SPECT reconstructed with IACT also demonstrated good accuracy (RDmax <10%) and reduced radiation dose.
- HCT-AC resulted in greater deviations (RDmax up to 19.76%) and fluctuating results with increased motion.
Conclusions:
- CACT and IACT-AC are effective in reducing artifacts and enhancing quantitation in myocardial perfusion SPECT/CT.
- IACT offers the additional benefit of reduced radiation exposure.
- These advanced AC methods provide superior diagnostic accuracy over traditional HCT-AC.