Assessment of a smartphone app (Capstesia) for measuring pulse pressure variation: agreement between two methods: A
Borja Barrachina1, Raquel Cobos, Noemi Mardones
1From the Anaesthesia and Resuscitation Unit, (BB); Araba Research Unit, Bioaraba Research Institute (RC) and Intensive Care Unit, (NM, AC, CV) Araba Integrated Health Organization (University Hospital), Vitoria-Gasteiz, Álava, Spain.
Background:
Less invasive and noninvasive methods are emerging for haemodynamic monitoring. Among them is Capstesia, a smartphone app that, from photographs of a patient monitor showing invasive arterial pressure, estimates advanced haemodynamic variables after digitising and analysing the pressure curves.
Objective:
The aim of this study was to compare the level of agreement between the analysis of the signals obtained from the patient monitor and a photograph of the same images using the Capstesia app.
Design:
Cross-sectional study.
Setting:
Araba University hospital (Txagorritxu), Vitoria-Gasteiz, Alava, Spain, from January to February 2015.
Patients:
Twenty patients (229 images) who had an arterial catheter (radial or femoral artery) inserted for haemodynamic monitoring.
Intervention:
Snapshots obtained from the patient monitor and a photograph of these same snapshots using the Capstesia application were assessed with the same software (MATLAB, Mathworks, Natick, Massachusetats, USA) for evaluating the level of concordance of the following variables: pulse pressure variation (PPV), cardiac output (CO) and maximum slope of the pressure curve (dP/dt). Comparison was made using interclass correlation coefficients with corresponding 95% confidence intervals, and Bland-Altman plots with the corresponding percentages of error.
Main Outcome Measures:
(PPV). Secondary outcome: CO and maximum slope of the pressure curve [dP/dt].
Results:
The interclass correlation coefficients for PPV, CO and max dP/dt were 0.991 (95% confidence interval 0.988 to 0.993), 0.966 (95% confidence interval 0.956 to 0.974) and 0.962 (95% confidence interval 0.950 to 0.970), respectively. In the Bland-Altman analysis, bias and limits of agreement of PPV were (0.50% ± 1.42) resulting in a percentage of error of 20% for PPV. For CO they were 0.19 ± 0.341, with a 13.8% of error. Finally bias and limits of agreement for max dP/dt were 1.33 ± 77.71, resulting in an error of 14.20% CONCLUSIONS: Photograph of the screenshots obtained with the Capstesia app show a good concordance with analysis of the original screenshots. Either approach could be used to monitor the haemodynamic variables assessed.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...