J Timothy Baldwin1, Iki Adachi2, John Teal3
1Division of Cardiovascular Sciences, National Heart, Lung and Blood Institute, Bethesda, MD.
Abstract:
The Infant Jarvik ventricular assist device (VAD; Jarvik Heart, Inc., New York, NY) has been developed to support the circulation of infants and children with advanced heart failure. The first version of the device was determined to have elevated hemolysis under certain conditions. The objective of this work was to determine appropriate modifications to the Infant Jarvik VAD that would result in acceptably low hemolysis levels. In vitro hemolysis testing revealed that hemolysis was related to the shape of the pump blade tips and a critical speed over which hemolysis would occur. Various design modifications were tested and a final design was selected that met the hemolysis performance goal. The new version was named the Jarvik 2015 VAD. Chronic in vivo tests, virtual fit studies, and a series of other performance tests were carried out to assess the device's performance characteristics. In vivo test results revealed acceptable hemolysis levels in a series of animals and virtual fit studies showed that the device would fit into children 8 kg and above, but could fit in smaller children as well. Additional FDA-required testing has been completed and all of the data are being submitted to the FDA so that a clinical trial of the Jarvik 2015 VAD can begin. Development of a Jarvik VAD for use in young children has been challenging for various reasons. However, with the hemolysis issue addressed in the Jarvik 2015 VAD, the device is well-poised for the start of the PumpKIN clinical trial in the near future.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation I: Indication and Settings
Motor Unit Stimulation
The latent period of contraction marks the onset of excitation-contraction coupling, when the action potential propagates across the sarcolemma, preparing the muscle fibers for contraction. As the fibers enter the contraction phase, the...
Cardiac Catheterization II: Right Heart Catheterization
Cardiopulmonary Resuscitation III: AED Use