Optimizing care of ventilated infants by improving weighing accuracy on incubator scales
1Neonatal Intensive Care Unit, Royal Bolton Hospital, Bolton, UK.
Objectives:
To determine the accuracy of weighing ventilated infants on incubator scales and whether the accuracy can be improved by the addition of a ventilator tube compensator (VTC) device to counterbalance the force exerted by the ventilator tubing.
Study Design:
Body weights on integral incubator scales were compared in ventilated infants (with and without a VTC), with body weights on standalone electronic scales (true weight). Individual and series of trend weights were obtained on the infants. The method of Bland and Altman was used to assess the introduced bias.
Results:
The study included 60 ventilated infants; 66% of them weighed <1000 g. A total of 102 paired-weight datasets for 30 infants undergoing conventional ventilation and 30 undergoing high frequency oscillator ventilation (HFOV) supported by a SensorMedics oscillator, (with and without a VTC) were obtained. The mean differences and (95% CI for the bias) between the integral and true scale weighing methods was 60.8 g (49.1 g to 72.5 g) without and -2.8 g (-8.9 g to 3.3 g) with a VTC in HFOV infants; 41.0 g (32.1 g to 50.0 g) without and -5.1 g (-9.3 g to -0.8 g) with a VTC for conventionally ventilated infants. Differences of greater than 2% were considered clinically relevant and occurred in 93.8% without and 20.8% with a VTC in HFOV infants and 81.5% without and 27.8% with VTC in conventionally ventilated infants.
Conclusions:
The use of the VTC device represents a substantial improvement on the current practice for weighing ventilated infants, particularly in the extreme preterm infants where an over- or underestimated weight can have important clinical implications for treatment. A large-scale clinical trial to validate these findings is needed.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.