Related Experiment Video
Updated: Mar 9, 2026

Fabrication of Electrochemical-DNA Biosensors for the Reagentless Detection of Nucleic Acids, Proteins and Small Molecules
Published on: June 1, 2011
An electrochemical DNA sensor without electrode pre-modification
Nian Hong1, Lin Cheng1, BingGuo Wei1
1Department of Pharmacy, JiangXi University of Traditional Chinese Medicine, JiangXi 330004, China.
Abstract:
We present a non-modification electrochemical DNA sensing strategy, which used Potential-Assisted Au-S Deposition and a clamp-like DNA probe. The dual-hairpin probe DNA was tagged with a methylene blue (MB) at the 3' terminal and a thiol at the 5' terminal., Without being hybridized with target DNA, the loop of probe prevented the thiol from reaching the bare gold electrode surface with an applied potential., After hybridization with the target DNA, the probe' s loop-stem structure opened through two distinct and sequential events, which led to the formation of a triplex DNA structure. Then the thiol easily contacted with electrode and resulted in potential-assisted Au-S self-assembly. Electrochemical signals of MB were measured by differential pulse voltammetry (DPV) and used for target quantitative detection. This strategy offered a detection limit down to 2.3pM. and an inherently high specificity for detecting even single mismatch.