Failure to detect extremely high levels of serum IgE with an immunoradiometric assay
J A Saryan1, P E Garrett, S R Kurtz
1Department of Allergy and Dermatology, Lahey Clinic Medical Center, Burlington, Massachusetts 01805.
Abstract:
Two patients with extreme elevations in serum IgE were found to have falsely low levels of IgE when measured by a commercial immunoradiometric assay. Dilution studies performed on serum from both patients revealed the rise and subsequent fall of the dose-response curve characteristic of the high-dose "hook" or prozone effect. Specimens with unexpectedly low levels of serum IgE obtained by either commercial immunoradiometric assay or enzyme immunoassay should be reanalyzed after dilution.