Contrast monitoring techniques in CT pulmonary angiography: An important and underappreciated contributor to breast
D P Mitchell1, M Rowan1, E Loughman1
1Department of Radiology, Whitty Building, North Circular Road, Mater Misericordiae University Hospital, Dublin 7, Ireland.
Objective:
The aims of our study were to evaluate the contribution of contrast-monitoring techniques to breast dose in pregnant and non-pregnant women, and to investigate the effect of a reduced peak kilovoltage (kV) monitoring scan protocol on breast dose and Computed Tomography Pulmonary Angiography (CTPA) diagnostic quality.
Materials And Methods:
Single center retrospective study of 221 female patients undergoing a reduced kV 80kV contrast-monitoring CTPA protocol compared to 281 patients using the conventional 120kV contrast-monitoring protocol (Siemens Somatom Definition AS+). 99 pregnant patients analyzed separately. ImPACT dosimetry software was used to calculate dose. Group subsets were evaluated to assess CTPA diagnostic quality.
Results:
The contrast-monitoring component of a CTPA study constituted 27% of the overall breast dose when using a standard 120kV protocol compared to only 7% of the overall breast dose in the 80kV study group. The dose to the breast from the contrast-monitoring component alone was reduced by 79% in the non-pregnant patients (0.36mGy±0.37 versus 1.7mGy±1.02; p<0.001), and by 88% in the pregnant population (0.25mGy±0.67 versus 2.24mGy±1.61; p<0.001). There was no statistical difference in CTPA diagnostic quality or timing.
Conclusion:
Despite a short scan length and relatively small DLP, contrast-monitoring techniques (test-bolus or bolus-tracked) set at 120kV can account for 27% of the overall breast dose accrued from a CTPA study. By decreasing the kilovoltage of the contrast-monitoring component, a significant reduction in breast dose for pregnant and non-pregnant female patients can be achieved without affecting CTPA quality or timing.
More Related Videos
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Pulmonary Embolism II: Diagnostic Studies and Interprofessional Care
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Pulmonary Embolism III: Nursing Management
Radiological Investigation I: X-ray and CT