Evaluation of optimised 3D turbo spin echo and gradient echo MR pulse sequences of the knee at 3T and 1.5T
O M Abdulaal1, L Rainford2, P J MacMahon3
1Diagnostic Radiology Technology, College of Applied Medical Sciences, Taibah University, Madina, Saudi Arabia; Radiography and Diagnostic Imaging, School of Medicine, University College Dublin, Dublin, Ireland.
Introduction:
To investigate the impact of parameter optimisation for novel three-dimensional 3D sequences at 1.5T and 3T on resultant image quality.
Methods:
Following institutional review board approval and acquisition of informed consent, MR phantom and knee joint imaging on healthy volunteers (n = 16) was performed with 1.5 and 3T MRI scanners, respectively incorporating 8- and 15-channel phased array knee radiofrequency coils. The MR phantom and healthy volunteers were prospectively scanned over a six-week period. Acquired sequences included standard two-dimensional (2D) turbo spin echo (TSE) and novel three-dimensional (3D) TSE PDW (SPACE) both with and without fat-suppression, and T2∗W gradient echo (TrueFISP) sequences. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were measured for knee anatomical structures. Two musculoskeletal radiologists evaluated anatomical structure visualisation and image quality. Quantitative and qualitative findings were investigated for differences using Friedman tests. Inter- and intra-observer agreements were determined with κ statistics.
Results:
Phantom and healthy volunteer images revealed higher SNR for sequences acquired at 3T (p-value <0.05). Generally, the qualitative findings ranked images acquired at 3T higher than corresponding images acquired at 1.5T (p < 0.05). 3D image data sets demonstrated less sensitivity to partial volume averaging artefact (PVA) compared to 2D sequences. Inter- and intra-observer agreements for evaluation across all sequences ranged from 0.61 to 0.79 and 0.71 to 0.92, respectively.
Conclusion:
Both 2D and 3D images demonstrated higher image quality at 3T than at 1.5T. Optimised 3D sequences performed better than the standard 2D PDW TSE sequence for contrast resolution between cartilage and joint fluid, with reduced PVA artefact.
Implications For Practice:
With rapid advances in MRI scanner technology, including hardware and software, the optimisation of 3D MR pulse sequences to reduce scan time while maintaining image quality, will improve diagnostic accuracy and patient management in musculoskeletal MRI.
Related Concept Videos
Magnetic Resonance Imaging
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies for Cardiovascular System IV: CMRI
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Imaging Studies III: Computed Tomography